Technical Data
Guidance and manufacturer's declaration – electromagnetic immunity
The SAM 500P is intended for use in the electromagnetic environment specified below. The customer or user of
the SAM 500P must assure that it is used in such an environment.
IEC 60601
Immunity test
test level
Conducted RF
IEC/EN 61000-4-6
150 kHz to 80 MHz
outside ISM bands
150 kHz to 80 MHz
in ISM bands
Radiated RF
IEC/EN 61000-4-3
80 MHz to 2.5 GHz
38
Compliance
level
3 Vrms
Not
applicable
a
Not
10 Vrms
applicable
a
10 V/m
10 V/m
80 MHz
to 2.5 GHz
Electromagnetic environment – guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the SAM
500P, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
Not applicable
Not applicable
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m)
.
b
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
should be less than the ... [next page]
,