b. If a child's chest is large enough to
permit a 1 inch (2.5cm) gap between the
pads, pad placement can be used similar
to adult placement. Place one pad on
child's BARE upper right chest above
nipple and one pad on child's BARE
lower left ribcage below nipple as shown
in Method b)
Method b)
Electrodes can be placed on the child's
chest if their chest is large enough OR if
trauma does not allow for the placement
as shown in Method a).
Warning: Defibrillation
electrodes must be at least
1 inch (approx 2.5 cm) apart
and should never be touching
one another.
Warning: The Pediatric-
Pak contains a magnetic
component (surface strength
6500 gauss). Avoid storage
next to magnetically sensitive
storage media.
Warning: Not for use on
patients under 1 year old. For
use with children up to the age
of 8 years or up to 55lbs (25kg).
DO NOT DELAY THERAPY
IF YOU ARE NOT SURE OF
EXACT AGE OR WEIGHT.
21