Technical Data

Physical parameters (with Pad-Pak installed)
Size:
Weight:
Environmental
Operating temperature:
Standby temperature:
Transport temperature:
Relative humidity:
Enclosure:
Altitude:
Shock:
Vibration:
20 x 18.4 x 4.8 cm (8.0 x 7.25 x 1.9 in)
1.1 kg (2.4 lbs)
0 to 50 °C (32 to 122 °F)
0 to 50 °C (32 to 122 °F)
-10 to 50 °C (14 to 122 °F) for up to two days. If the device
has been stored below 0 °C (32 °F), it should be returned to
an ambient temperature of between 0 to 50 °C (32 to 122 °F)
for at least 24 hours before use.
5 to 95% (non-condensing)
IEC 60529/EN 60529 IP56
0 to 15 000 feet (0 to 4575 metres)
MIL STD 810F Method 516.5, Procedure 1 (40G's)
MIL STD 810F Method 514.5 Procedure 1 Category 4
MIL STD 810F Method 514.5 Procedure 1 Category 7