4. Performing the DICOM conformance test
Press either + or - button to select "Yes" for START TESTING. (DICOM conformance test starts immediately).
It takes between 30 seconds to 1 minute to complete the test.
To cancel the test, press either + or - button to select "Yes" for CANCEL TESTING.
5. Showing the test results
When the test has completed, the results are shown similar as below.
Luminance status
The result shown indicates the following status.
RESULT
Passed
Failed
Canceled
Passed(!)
Failed(!)
LUMINANCE
Stable
Unstable
NOTE:
If the test is repeated within 30 seconds, "Unstable" is shown for LUMINANCE and "(!)" is show n for RESULT.
The test results are saved until the next test or calibration.
If the maximum deviation (GSDF error) result does not conform to the AAPM TG18 guidelines, even though the test has been
performed under the recommended conditions, calibrate the display using an external calibration sensor, i.e. the NEC model
MDSVSENSOR3. During calibration with an external sensor, the front sensor of the display will be re-adjusted as well.
NOTE:
The OSD menu disappears in 15 seconds.
Screen that shows the test is in execution
Test result
Maximun deviation
Elapsed time
Start of test
Max. deviation (GSDF error) ≤ 10%, complies with AAPM TG18 guidelines for
primary displays.
Max. deviation (GSDF error) > 10%, does not comply with AAPM TG18 guidelines
for primary displays.
Test has been cancelled.
When "Passed" or "Failed" is followed by "(!)," the test has not been performed
under the recommended condition (minimum 60 minutes of active backlight time)
or the luminance is unstable.
Luminance is stable.
Luminance is unstable.
Test result
Explanation
Explanation