Intended Use

The MD302C6 color monitor is intended to be used for displaying and viewing of digital images by trained physicians.
Caution: The MD302C6 color display cannot be used for a life-support system.
This device is designed as component of a fi nal system which is in compliance to IEC 60601-1-1 requirements.
Contraindications: None known.
Safety Precautions
This device is designed for exclusive interconnection with IEC 60950-1 certifi ed equipment outside of patient vicinity.
• Equipment connected to digital interfaces must comply with the respective IEC standards (e.g. IEC 60950-1 for data
processing equipment and IEC 60601-1 for medical equipment).
• This device complies with IEC 60601-1-2. To minimize the interference from other equipment, a minimum 0.5 m distance shall
be kept form other potential electromagnetic sources, such as a Cell Phone.
• To reduce the risk of electric shock, make sure that the power cord is unplugged from wall socket. To fully disengage the
power to the unit, please disconnect the power cord from the AC outlet. Do not open the back cover. No user serviceable parts
inside. Refer servicing to qualifi ed service personnel. The AC outlet shall be readily available and accessible.
EMC Precaution
This product requires special EMC precautions and needs to be installed and used according to the following information.
Use the specifi ed cable. Using other cables could increase emission or decrease immunity.
Do not place any RF communication equipment in the vicinity of this product. It could affect the performance of the product.
Any additional equipment connected to this product needs to comply with the requirements of IEC 60601-1-2.
North America Customers
Grounding reliability can only be achieved when the equipment is connected to an equivalent receptacle, marked "Hospital Only"
or "Hospital Grade". These signal input ports need to be connected properly and any unused signal input ports shall not be
accessible for the patient to contact in the patient area after the LCD is integrated into a medical system.
Federal law restricts this device to sale by or on the order of a licensed healthcare practitioner.
European Customers
Unpacking, Installation and calibration of this display should only be done by authorized and trained personnel. Any installation
done by a non-authorized person is done under his own risk and we accept no responsibility in any device malfunctioning. To
secure full functionality of this display, only NEC approved display controller models for connection with the fi nal system must be
used.
Medical Imaging
MD302C6 is designed for 3280 x 2048 medical imaging.
English-4