Respironics HomeLOX User Manual page 8

Table of Contents

Advertisement

C
p
onTrol
anel
Description
SERVICE
Indicator
1
PAUSE
Push Button &
Indicator
2
START
Push Button &
3
Indicator
LIQUID OXYGEN
SEGMENT LEVEL
Indicators
4
HomeLox User Manual
3
2
1
Color
Function
When lit, this indicates the
unit has detected a fault
that requires service by the
Yellow
equipment provider. The
unit has shut itself off and
will not start again until the
problem is corrected.
When this switch is
depressed, the unit stops
Blue
operation for a period of 1
hour after which operation is
resumed.
When this switch is
depressed the unit goes
Green
into its operating state to
produce liquid oxygen.
When each segment is lit, it
indicates that the GoLox-93
can be filled once.
When the first segment
Blue
(next to E) is lit, you can fill
your GoLox-93 once.
When the gauge is at F, with
all six segments lit, you can
fill your GoLox-93 six times.
4
Indicator
Modes
Steady On – The HomeLox has detected a problem that
prevents it from operating. Contact your home care
equipment provider for repair.
Off – No system faults have been detected.
Steady On – The HomeLox is paused; it is in a temporary non-
operating state.
Off – The HomeLox is not in PAUSE mode.
Steady On – Unit is turned on; mains power is connected.
Flashing – Unit is operating to produce liquid oxygen.
Off – The HomeLox is turned off; mains power is not
connected.
Segment Steady On – Each lit segment is equivalent to 1 fill
of your GoLox-93 from the HomeLox. The total number of lit
segments is equal to the total number of times your GoLox-93
can be filled.
One Segment Flashing – This segment of the HomeLox
storage bottle is being filled.
Segment Off – This segment of the HomeLox storage bottle
has not yet been filled.
7