Carephone 62
63 Acoustical feedback for technical failures
The Carephone 62 can be set to announce technical failures through the
LED lamps and the loudspeaker, or through the LED lamps only.
0 = loudspeaker off
1 = loudspeaker on
2 = loudspeaker on from 7:00 to 21:00
3 = an acoustical failure indication (message or beep) shall be
repeated with a pause of 10 seconds until the Action button is
pressed.
4 = an acoustical failure indication (message or beep) shall be
repeated during day time (from 7:00 to 21:00) with a pause of 10
seconds until the Action button is pressed.
Settings 2 and 4 are only available if the time and date are set. This can be
done through keyboard programming, Configuration Manager with serial
connection or via a remote date/time update from a receiver.
70 Automatic test call interval
You can program the interval between automatic test calls from the
Carephone 62 to the monitoring centre in days. In addition, there is a test
call after power up or leaving programming mode, as well as a test call per
randomization.
Select from 1 to 28 days.
0 = off
Default value is 0.
71 Action button (S button)
The Action button can be configured differently according to your needs:
0 = off
1 = service button / direct call (e.g. service call to a monitoring centre
or direct call to a relative)
3 = activate relay output (e.g. a door opener)
4 = send a manual test call (by pressing at least 2 seconds on the
action button)
Default value is 0.