Specifications - Sanitas SBM 50 Instructions For Use Manual

Hide thumbs Also See for SBM 50:
Table of Contents

Advertisement

Available languages

Available languages

11. specifications

Model no.
SBM 50
Measurement
Oscillometric, non-invasive blood
method
pressure measurement on the up-
per arm
Measurement range
Cuff pressure 0 – 300 mmHg,
systolic 60 – 260 mmHg,
diastolic 40 –199 mmHg,
Pulse 40 –180 beats/minute
Display accuracy
Systolic ± 3 mmHg,
diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement
Max. permissible standard deviation
inaccuracy
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
Memory
2 x 60 memory spaces
Dimensions
L 100 mm x W 150 mm x H 58 mm
Weight
Approx. 475 g (without batteries)
Cuff size
22 to 30 cm
Permissible operating
+5 °C to +40 °C, ≤90 % relative air
conditions
humidity (non-condensing)
Permissible storage
conditions
Power supply
Battery life
Accessories
Classification
Technical information is subject to change without notifica-
tion to allow for updates.
• This unit is in line with European Standard EN 60601-1-2
and is subject to particular precautions with regard to
electromagnetic compatibility (EMC). Please note that
portable and mobile HF communication systems may in-
terfere with this unit. More details can be requested from
the stated Customer Service address or found at the end
of the instructions for use.
• This device is in line with the EU Medical Devices Direc-
tive 93/42/EC, the "Medizinproduktegesetz" (German
Medical Devices Act) and the standards EN 1060-1 (non-
28
-20 °C to +55 °C, ≤95 % relative air
humidity, 800 –1050 hPa ambient
pressure
4 x 1,5 V
AA batteries
For approx. 500 measurements, de-
pending on the blood pressure level
and/or pump pressure
Instruction for use, 4 x 1.5 V AA bat-
teries, storage pouch
Internal supply, IPX0, no AP or APG,
continuous operation, type BF ap-
plied part

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents