Corrective Action - Respironics Millennium H600 Service And Technical Reference Manual

Respironics millennium oxygen concentrators
Table of Contents

Advertisement

SYMPTOM
Unit will not alarm
(Red LED)
Power Cord is
connected
Power Switch in the
ON position
No Audible Alert
No Red LED
LED failure (Models
600 and 605)
Red LED
inoperable
Green LED
inoperable
Yellow LED
inoperable
P
2 - T
AGE
ROUBLESHOOTING
CAUSE
No power to the
alarm
No power to LEDs
M
S
ILLENNIUM
ERVICE
VERIFICATION
Verify the battery
voltage is 5 volts
or higher
Verify the battery
connector is
correctly seated
on the battery
Verify the wiring
harness is
connected to the
Power Switch
Verify the wiring
harness is
connected to the
Main PCB
Verify the Power
Cord is plugged in
Verify the battery
voltage is greater
than 5 volts
Verify
that
all
wiring harnesses
are connected
Verify
the
Main
PCB
is
not
damaged
Verify
the
OPI
board
is
not
damaged (Model
605 only)
& T
I
ECHNICAL
NFORMATION

CORRECTIVE ACTION

Replace the 9 volt battery
Connect the wiring
harness
Connect the wiring
harness
Connect the wiring
harness
Replace the Main PCB
Connect the Power Cord
to AC power source
Replace the 9 volt battery
Connect
all
harnesses
Replace the Main PCB
Replace the OPI board
1041814, R
wiring
. 00
EV