Sheet - Respironics Millennium H600 Service And Technical Reference Manual

Respironics millennium oxygen concentrators
Table of Contents

Advertisement

M
600
/P
ODEL
W
EDIATRIC
All information on this data sheet should be entered in the correct location after the associated test was
completed. The data sheet must then be signed in ink and dated by the technician performing tests.
Step 2
Serial Number: __________
Step 5 is not required for this Model. The Alarm will not
sound with the 1 Liter Pediatric Flowmeter installed.
Step 6
Oxygen Outlet Pressure
5.0-7.0 psig
NOTE
Step 7 is not required for this Model.
Step 8
Flow Meter
Oxygen
Concentration
Signature: ____________________ Date:_______________ Notification Number: _____________
1041903, R
. 00
EV
F
M
LOW
ETER
NOTE
Step 3
VAC: __________
NOTE
PRESSURE GAUGE READING
H600; M600; Enhanced M600
120V
> 92% @ 0.1 lpm to 1 lpm
0.1 lpm
1 lpm
M
S
ILLENNIUM
ERVICE
T
D
S
EST
ATA
HEET
Hz:__________
& T
I
ECHNICAL
NFORMATION
Step 4
Hour Meter: __________
(RI internal use Only)
T
- P
ESTING
AGE
11