Respironics Millennium H600 Service And Technical Reference Manual page 315

Respironics millennium oxygen concentrators
Table of Contents

Advertisement

M
605
/P
ODEL
W
EDIATRIC
All information on this data sheet should be entered in the correct location after the associated test was
completed. The data sheet must then be signed in ink and dated by the technician performing tests.
Step 2
Serial Number: __________
Step 5 is not required for this Model. The Alarm will not
sound with the 1 Liter Pediatric Flowmeter installed.
Step 6
Oxygen Outlet Pressure
5.0-7.0 psig
Step 7
Oxygen
1ST CYCLE
Alalyzer
Reading
Average O2 Purity (3 cycles)
DMM Voltage Reading
PASS
Step 8
Flow Meter
Oxygen
Concentration
Signature: ____________________ Date:_______________
P
10 - T
AGE
ESTING
F
M
LOW
ETER
NOTE
Step 3
VAC: __________
NOTE
PRESSURE GAUGE READING
2ND CYCLE
FAIL
H605; M605; Enhanced M605
120V
> 92% @ 0.1 lpm to 1 lpm
0.1 lpm
M
S
ILLENNIUM
ERVICE
T
D
S
EST
ATA
HEET
Hz:__________
3RD CYCLE
1 lpm
Notification Number: _____________
& T
I
ECHNICAL
NFORMATION
Step 4
Hour Meter: __________
(RI internal use Only)
1041903, R
. 00
EV

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents