MESSAGE
58 mg/dL
(Example)
Anotok
redo A
retest
PROBLEM
The check strip test result is
outside the acceptable range
(printed on the back of
the meter).
The last check strip test was
outside the acceptable range
and an acceptable repeat
test was not performed.
The last test failed after the
countdown started.
WHAT TO DO
Clean the meter and check
strip if necessary. Repeat the
check strip test.
Repeat the check
strip test.
Repeat the test with a
new test strip.
70