MESSAGE
100 mg/dL
(Example)
contrl
error 1
retest
error2
retest
PROBLEM
If the word contrl appears after
a blood test, your blood sample
was too small, smeared, or
another drop was added after
the test began.
The sample was applied before
the words apply sample
appeared.
1. The test strip moved during
the test.
2. The test strip was not inserted
correctly.
WHAT TO DO
Repeat the test with a new
test strip and a larger drop of
blood.
Repeat the test with a new
test strip.
1. Repeat the test with a
2. Repeat the test with a
65
new test strip.
new test strip.