Printing Problems; Status Led; Solutions - Midtronics PowerSensor Micro700 Instruction Manual

Heavy duty electrical system analyzer
Table of Contents

Advertisement

PRINTING PROBLEMS

Status LED

When a printer fault occurs, the STATUS LED will flash. You can
identify the fault by the number of sequential flashes:
Sequence
* * *
** ** **
*** *** ***

Solutions

If the IR transmitter and receiver are not aligned, all the data may
not print. The infrared ports on the top of the Micro700 and on
the front of the printer below the MODE button should be pointed
directly at each other. The maximum distance for reliable
transmission between the ports is 18 inches (45 cm). To realign,
press MENU to cancel the print job. Verify alignment between
the analyzer and printer; then try to print the test results again.
If the analyzer displays PRINTING RESULTS, but the test results
are not printing, press MENU to cancel. Turn off the printer and
charge the printer battery for at least 15 minutes before
attempting to print again.
Make sure the printer is on. The printer turns itself off after 2
minutes of inactivity to conserve the battery. To turn the printer
on, briefly press the MODE button. The green STATUS light
should turn on. Make sure you are using the Midtronics printer
provided with the Micro700. Other printers may not be
compatible.
Direct sunlight interferes with infrared data transmission/
receiving. If the printer is not receiving data, remove the printer
and Micro700 from direct sunlight. If the printed characters are
not clear or are partially missing, recharge the battery and
reprint.
Condition
Paper out
Thermal head too hot
Battery low
17
Solution
Fit new paper
Allow head to cool
Recharge printer
battery for 16 hours

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents