56
Section 7
Calibrate Your System/Test Blood Glucose Manually
Steps
1.
Insert a FreeStyle Test Strip into
the receiver's test strip port.
2.
Ensure that the code number
on the screen matches the code
number on the test strip vial.
If
• The numbers match. a. Press the RIGHT Option button (Set).
• The numbers do
NOT match.
• The screen changes
to the Apply Sample
screen.
Customer Care: 1-866-597-5520
Do This
a. Insert the end with the dark rectangle.
b. Gently push until the test strip stops.
Result: The receiver will display the most recently used
test strip code number.
Caution: The code numbers MUST match to ensure
accurate test results.
Compare the code number on the display with the code
number on the vial of test strips.
Then
b. Go to Step 3.
a. Use the UP/DOWN Arrow buttons to change the
code to match.
b. Press the RIGHT Option button (Set).
c. Go to Step 3.
If the numbers do NOT match:
a. Press the LEFT Option button (Back).
b. Repeat Step 2.
If the numbers match, go to Step 3.
Result: The receiver will prompt you to "Apply Sample. "