Section 18
Appendix B: Specifications
18 Appendix B: Specifications
Receiver Size
Receiver Weight
Receiver Battery Life
Automatic Shutoff
Receiver Memory
Calibration
Calibration Time
Blood Glucose Test Time
(for BG mode)
Blood Glucose Assay Method
(for BG mode)
Blood Sample Type (for BG mode)
Hematocrit (for BG mode)
Height: 2.5 in. (6.3 cm). Width: 3.24 in. (8.2 cm).
Depth: 0.88 in. (2.2 cm).
3.5 oz. (99.2 grams) – including batteries.
Up to 60 days.
Built-in blood glucose meter: 2 minutes after last user
action. Receiver: 12 seconds after last user action.
• 60 days of normal use including continuous glucose
readings (stored every 10 minutes) and daily blood
glucose readings.
• Date/time will be remembered for 5 minutes after receiver
battery removal.
Plasma equivalent.
• 1st calibration: Must be performed at approximately
10 hours after a new sensor has been inserted. The
first calibration can be performed after the 10 hours.
Continuous glucose readings will NOT be reported until
the 1st calibration is performed successfully.
• 2nd calibration: Must be performed between 2 and 4 hours
after the 1st calibration or continuous glucose will not
be reported. The 2nd calibration can be performed after
4 hours; continuous glucose reporting will resume after
completing the 2nd calibration successfully.
• 3rd calibration: Must be performed between 12 and 20
hours after the 2nd calibration or continuous glucose will
not be reported. The 3rd calibration can be performed
after 20 hours; continuous glucose reporting will resume
after completing the 3rd calibration successfully.
• 4th calibration: Must be performed between 48 and 56
hours after the third calibration or continuous glucose will
not be reported. The fourth calibration can be performed
after 56 hours; continuous glucose reporting will resume
after completing the 4th calibration successfully.
Average of 7 seconds.
Coulometric electrochemical sensor.
Whole blood, capillary.
15% to 65%.