When you reach the last test result, the word "End" will appear,
and the meter will shut o automatically. Similarly, the meter will
shut o automatically when it is idle for three minutes.

CLEANING AND DISINFECTION

Cleaning and disinfection is a necessary and important part of
the test procedure. It can help to prevent infection, the potential
spread of infection, and cross-contamination. Cleaning can also
ensure that the meter works properly and that the display is
clear and readable.
The meter and lancing device should be cleaned and disinfected
following each use. We suggest that you use CaviWipes™
(Metrex® Research Corporation, EPA Reg. No. 46781-8, EPA Est.
No. 56952-WI-001). CaviWipes, with isopropanol and
diisobutyl-phenoxy-ethoxyethyl dimethyl benzyl ammonium
chloride as the active ingredient, have been shown to be safe for
use with the meter and lancing device.
You can purchase this product from the suppliers listed below:
1 Visit the website www.metrex.com or contact Metrex at
800-841-1428 for product or technical information.
2 Visit the website
http://www.endochoice.com/Equipment?search=wipe.
The meter and lancing device are validated to support 10,000
individual tests—and consequently 10,000 cleanings over their 5
year life spans.
Below are the steps on how to clean the meter and lancing
device.