Non Pressurized Demineralized Water Connection - Smeg GW2045 Operating Instructions Manual

Glassware washer
Table of Contents

Advertisement

GW1060
fig. 5 - rear of product GW1060, detail of rear cross-section with indication of water inlets and outlets The appliance is standard equipped with 1

5.4. NON PRESSURIZED DEMINERALIZED WATER CONNECTION

The connection to a non-pressurized demineralized water supply (e.g. gravity tank) is possible only by
installing the special demineralized water pump, called "SMEG-PAD".
This pump makes sure that the demineralised water enters at sufficient pressure to make the appliance work
correctly. The PAD accessory can be fitted externally on any machine and must be mounted on the back of
the machine.
We advise having the PAD installed by a specialized SMEG technician.
peristaltic pump for detergent dosage: P2. P1 and P3 are optional accessories.
If the PAD pump is used, the water supply pressure must be less than 1 bar.
WARNING – WHEN INSTALLING THE PAD, MAKE SURE THAT:
Make sure that the demineralized water pressure is less than 1bar. Smeg declines all
responsibility if the PAD is installed with the inlet water at a higher pressure.
Smeg will accept no responsibility for any possible damage to items or people
caused by an incorrect installation performed by an unauthorized technician.
Rev. 00
GW2045 – GW1060 Manual
Pag. 16 di 43