ALARM ID
The cold water flowmeter does not
22
measure correctly.
23
Insufficient water.
24
Lack of water.
Insufficient pressure in the circulating
25
motor pump.
29
No draining.
30
Water safety level.
31
Safety level fault.
Standing water in the washing
32
chamber.
33
OFF
34
OFF
35
OFF
36
OFF
52
Door is electrically open.
54
Door is mechanically open.
56
Door lock fault.
73
Cycles archive is full.
Inlet water temperature higher than
77
45° C, prewashing must be carried
out at temperatures below 45° C.
78
Microprocessor failure.
Demineralised water inlet problem –
81
not active.
82
Door lock coil fault.
94
Unstable temperature
Rev. 00
DESCRIPTION
GW2045 – GW1060 Manual
Check the water supply (taps, pressure, connections to
pipes etc), repeat programme.
Check the water supply (taps, pressure, connections to
pipes etc).
Check the water supply (taps, pressure, connections to
pipes etc).
Possible presence of foam in vat.
Check the type of detergent used.
Repeat the cycle.
Check the outlet connection.
Repeat the cycle. Call the Technical Assistance Service if
the fault persists.
Repeat the cycle. Call the Technical Assistance Service if
the fault persists.
Reset. Repeat the cycle. Call the Technical Assistance
Service if the fault persists.
Check that the door is closed before proceeding with a
programme cycle.
Open and close the door. Check the condition of the
micro-switch that detects closing. Call the Technical
Assistance Service if the fault persists.
Check that the door is closed before proceeding with a
programme cycle. Call the Technical Assistance Service
if the fault persists.
Check that the door is closed before proceeding with a
programme cycle. Call the Technical Assistance Service
if the fault persists.
Error within the microprocessor. These are overwritten
one at a time once the cycles archive becomes full.
Call the Technical Assistance Service if the fault persists.
The alarm trips when the temperature at the start of the
cycle is greater than 45° C.
Wait for the machine to cool before starting a new cycle.
Error within the microprocessor.
Call the Technical Assistance Service if the fault persists.
Check the water supply:
Tap open
!
Inlet water pressure
!
Absence of narrowing in connection pipe).
!
Check the condition of the demineralised water
!
turbine.
Repeat the cycle, the malfunction may be temporary.
Repeat opening. Call the Technical Assistance Service if
the fault persists.
The active alarm in the door opening phase: If, after a
series of opening impulses, the door remains closed,
intervene. Check that the solenoid run is not blocked and
that the opening piston on the side of the vat releases
correctly.
This malfunction can be temporary: Repeat the cycle.
Check that the temperature probe is functioning properly.
ACTION