Page 1
IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODELS: 61-732 61-736 The Service Information provides the following information: • Precautions and safety information • Specifications • Performance test procedure • Calibration and calibration adjustment procedure • Basic maintenance (replacing the battery) HOLD Ω...
Page 2
TABLE OF CONTENTS Page Title Introduction Precautions and Safety Information Symbol Table Safety Information Certifications and Compliance Specifications General Specification Voltage Specifications Current Specifications Resistance Specifications Continuity Specifications Performance Verification Calibration Calibration Reference Points Replacing the Batteries Form number TM61732-6 Rev 5 October 2004...
Page 3
The 61-730 series contains parts that can be damaged by static discharge. Follow the standard practices for handling static sensitive devices. For additional information about IDEAL INDUSTRIES, INC. and its products, and services, visit IDEAL INDUSTRIES, INC. web site at: www.idealindustries.com Precautions and Safety Information Use the meter only as described in the Users Manual.
Page 4
Page 2 SAFETY Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use the product only as specified. CAUTION. These statements identify conditions or practices that could result in damage to the equipment or other property.
Page 5
-20°C to 60°C (-4°F to 140°F) at <80% relative humidity Power source: 1.5V Battery x 2 (R03/Size AAA) Battery Life: 400 hours typical (alkaline) {61-732} 250 hours typical (alkaline) {61-736} Low Battery Indicator: symbol indicates low battery voltage Auto power off...
Page 6
Not specified Continuity <400Ω on Continuity AC Converter: 61-732 is Average responding, RMS Calibrated to Sine Wave. 61-736 is True RMS responding. Overload Protection: AC and DC Voltage: Not to exceed 600V DC or VAC RMS. AC Current: Not to exceed 400A AC.
Page 7
Table 2 DC Voltage Test Function High Input Model Number /Range Limit Limit V DC 400V 350V 348.0 352.0 61-732, 61-736 V DC 600V 500V 61-732, 61-736 Table 3 AC Current Test Function High Input Model Number /Range Limit Limit...
Page 6 Table 4 Resistance Test Function High Input Model number /Range Limit Limit Ω 400 98.6 101.4 61-732, 61-736 100Ω Ω 4K .986 1.014 61-732, 61-736 1KΩ Ω 40K 9.86 10.14 61-732, 61-736 10KΩ Ω 400K 98.6 101.4 61-732, 61-736 100KΩ...
Page 9
De-energize source and remove test leads. A AC Calibration: 61-732 (Refer to Figure 1B), 61-736 (Refer to Figure 2B) (Adjustments made under front panel label.) Set the function / range to the 400A AC. Set output of the AC calibrator for 10.00A/60Hz +/- 0.01% and connect it to Coil = 10N = 100.0A Clamp the jaws to the coil = 10N.
Page 10
Page 8 Figure 1A (61-732) Figure 1B (61-732) Figure 2A (61-736) Figure 2B (61-736) Form number TM61732-6 Rev 5 October 2004...
Page 11
Page 9 Battery Replacement (refer to Figure 3) 1. Disconnect the test leads from any circuit under test and turn off meter. 2. Use a Philips head screwdriver to remove the screws on battery cover. 3. Remove batteries from the battery compartment. 4.