15.6 Calibration with external weight (CALIb)
Calibration with external weight standard in verified balances should be performed in case balance indications
exceed permissible error. To calibrate the balance a service centre should use calibration weight as stated in
Technical Data table (or of better accuracy).
Operations sequence:
m
0.00g
MENU
CALIbr
T
CAL oFF
...
CAL on
T
- CAL -
0.00g
LOAD
m
0.00g
Klawiszem MENU wywołać pojawiające się kolejno
funkcje użytkownika.
Press MENU key to call functions menu and to
→
choose CALIbr with
T
Calibration function options:
- CAL oFF – turn autocalibration off
- CAL on – start calibration with external weight
- out – exit without saving
→
←
Press press
to choose CAL on.
T
Wait until zeroing is finished.
When LOAD communicate put a calibration weight on
the pan.
Wait until calibration process is finished and zero
indicated.
info@tursdaletechnicalservices.co.uk
←
key.
16