info@tursdaletechnicalservices.co.uk 1. Introduction Electrical balance BT series are designed for use in laboratories and many others technical applications, where high measurements precision is not required. All balances are metrologically tested by manufacturer and they are not designed for calibration certificate. According to order, balances may be calibrated.
info@tursdaletechnicalservices.co.uk 6. Keys and indicators Keys functions description: - switch-on/switch-off (standby), - numeric key / menu, MENU → ← - tare (subtract package weight from weighed mass) - numeric key / data output (print / transmission), → ← - numeric key / zeroing (balances for direct sale use only), indicator - result stabilisation, bar indicator...
Page 6
info@tursdaletechnicalservices.co.uk +33°C +18°C 8. Balance preparation for work 1. Remove balance and power supply unit from the package. 2. Place the balance on stable foundation, in place not subjected to mechanical vibrations and air movements. 3. Unscrew protection rod 10 and take it out with protection cap 11 (if it appears in scale).
Page 7
info@tursdaletechnicalservices.co.uk Attention: Pos. 10 and 11 are optional. If balance was moved from environment with lower temperature to the room with higher temperature, condensate can be formed on the surface of balance. In this situation it is forbidden to turn balance on, because balance can be damaged. Before turning on balance, it should stay in a new environment for 4 hours in order to acclimatization.
Page 8
info@tursdaletechnicalservices.co.uk 3. Charging of accumulators is made automatically, when balance is connected to power supply unit (also during weighing) Charged state of accumulators can be checked by auxiliary function bAttErY (bat VoL option) 11. Accumulators (batteries) exchange 1. Remove pan lid that shields scale pan. Remove pan cover and put it near the balance.
info@tursdaletechnicalservices.co.uk 12. Start-up Leave the pan empty, plug the feeder to the mains (~230V/50Hz) and plug the feeder connector to the 12V power socket at the back of the scale The scale proceeds with following start-up actions: Basic electronic components tests: C1, C-2, ... C-6. bt...
Page 10
info@tursdaletechnicalservices.co.uk Standard transmission parameters: 8bits, 1stop, no parity, 4800bps. To change transmission parameters use "rS-232" function (see further part of the manual). Data transmission protocol description (Long protocol) Computer→Balance: initialising signal S I CR LF (53h 49h 0Dh 0Ah) Balance→Computer: balance indication in the following format (16Bytes) Byte sign or space Byte...
Page 11
info@tursdaletechnicalservices.co.uk 14. Basic balance functions In further part of this manual the following graphical symbols will be used for balance functions description. put load on the pan put load of the pan press a key during indication forced change automatic change 14.1 Normal weighing When the pan is empty and indication is different than 0.00 g...
Page 12
info@tursdaletechnicalservices.co.uk 15. Balance special functions All balances, besides the basic metrological functions: weighing and tare setting, have a set of balance special functions. Basic set of includes the following special functions: - Automatic zeroing function (AutotAt) - pieces counting (PCS), - Convert mass units (UnIt) - Convert percentage (PErCEnt) - Preparation of recipes (rECIPE)
Page 13
info@tursdaletechnicalservices.co.uk 15.2 Pieces counting function (PCS) That function allows for counting the identical details, e.g. tablets or buttons, contained in the weighed batch. 12.05 g The measurement takes place in two phases: - first phase – calculation of single detail weight upon the basis of sample of specified number of pieces 5, 10, 20, 50, 100, 200, or 500 detils, 0.00 g...
Page 14
info@tursdaletechnicalservices.co.uk 15.3 Weighing unit selection (UnIt) Use this function to choose weighing unit: 1.60g MENU CarAt (1 ct= 0,2 g), MGrAM (1mg=0,001g), KGrAM (1kg=1000g), UnIt Pound (1 lb=453,592374g), Ounce (1oz=28,349523g), OunCEt (1ozt=31,1034763g), GrAIn (1gr=0,06479891g) CArAt PennYW (1dwt=1,55517384g), Gram (1g). The example at the left presents how to set carats as weighing unit.
Page 15
info@tursdaletechnicalservices.co.uk 15.5 Weigh summing (rECIPE) This function enables to weigh few ingredients in one 0.00 g container and to display aggregated sum of all MENU ingredients. The function has the following options: rECIPE -rEC oFF – leave the function and display aggregated sum of all ingredients, -rEC on –...
Page 16
info@tursdaletechnicalservices.co.uk 15.6 Calibration with external weight (CALIb) Calibration with external weight standard in verified balances should be performed in case balance indications exceed permissible error. To calibrate the balance a service centre should use calibration weight as stated in Technical Data table (or of better accuracy). Operations sequence: Klawiszem MENU wywołać...
Page 17
info@tursdaletechnicalservices.co.uk 15.7 Serial port setting (rS-232) The function enables to set the following transmission parameters: 0.00g MENU - bAud (4800, 9600, ..., 115200) - transmission speed, - bitS (7, 8) - the number of bits in a byte, rS -232 - PArItY (0, 1) - parity control, - Odd (0, 1) –...
Page 18
info@tursdaletechnicalservices.co.uk Charging non-rechargeable batteries to test can cause damage of balance. 15.9 Display setting function (b_LIGHt) The function is used to choose action method of display 0.00g MENU bright light b_L OFF – bright light function deactivation, b_L on – bright light function activation, b_LIGHt b_L ECO –...
Page 19
info@tursdaletechnicalservices.co.uk 15.11 Reading unit choosing (rESOLUt) 0.00g MENU This function enables to choose reading unit value (resolution). The function is only available in BT200 and rES OLUt BT2000. Available reading units: wagi BT200: d=0,01g i d=0,001g, wagi BT2000: d=0,1g i d=0,01g. 0.000g 15.12 Function Menu customisation (ACtIV i dEFAULt) 0.00...
info@tursdaletechnicalservices.co.uk 16. Maintenance and repairs of small defects 1. The balance should be kept clean. 2. Take care that no dirt gets between the casing and the pan. If found any, remove the pan (lift it up), remove dirt and then replace the pan. 3.
info@tursdaletechnicalservices.co.uk Declaration of Conformity AXIS Spółka z o.o. 80-125 Gdańsk, ul. Kartuska 375B confirm with all responsibility that BT series balances: BT200, BT200D, BT2000 i BT2000D marked with CE mark comply the following: 1. EN 55022 standard Limits and methods of measurement of radio disturbance characteristics of information technology equipment and IEC 61000-4-3 Electromagnetic compatibility (EMC) - Part 4-3: Testing and measurement techniques - Radiated, radio-frequency, electromagnetic field immunity test,...