info@tursdaletechnicalservices.co.uk 1. General description Drive-through series in basic version from lacquered steel, with SE-01/A/18 meter, are designed for operation in light environmental conditions. Manual describes standard balance version. In case of special meter use, e.g. dosing meter, it is necessary to use documentation for meter as well. All balances are tested in respect of metrology.
info@tursdaletechnicalservices.co.uk 4. Safety principles is necessary to become familiar with safety principles shown below, which are necessary to avoid electric shock and damage of balance or connected devices. Repairs and necessary adjustments must be performed by qualified personnel only. To avoid fire hazard, use only typical supply cable, and supply voltage must be consistent with technical data.
info@tursdaletechnicalservices.co.uk 6. Balance keys and indicators ON/OFF switch (standby), tare setting (entering of mass subtracted from weighed mass), gross weight indication, zeroing, MENU special functions menu, result print, weight indication resolution increasing, indicator zero indicator (for unloaded balance), indicator signals weighing result stabilization, indicator net weight (after T key use),...
info@tursdaletechnicalservices.co.uk 7. Balance preparation to work 1. Unpack balance, removing protective foils. 2. Place balance flat, horizontal foundation, place subjected mechanical vibrations and strong movements. Take care if the scale stays on all four legs. 3. Air bubble in level should be located in the middle position.
info@tursdaletechnicalservices.co.uk 8. Balance start Connecting of supply plug to socket of ~230V/50Hz installation with unloaded balance will cause the following sequence of actions: Display test Tests of basic electronic subassemblies, ended with messages: C1, C-2, ... C-6. Displaying of software version. Ready for operation.
info@tursdaletechnicalservices.co.uk the result stabilization. 5. When there is no weighing, but the balance should be ready for operation, it may be switched off by I/ key. It causes the balance reading system deactivation end entering the standby mode. Balance turning on is performed by pressing "I/ " key. 6.
info@tursdaletechnicalservices.co.uk 12. Connection with computer or printer (option) For data transfer to external devices the balance is equipped with RS232C or RS485 (option) connector. balance sends the weighing result after the initiating signal from the computer, pressing key. During the operation with printer data are sent automatically after sampling and reading stabilization, and the next transmission is possible after the sample removing.
Page 11
info@tursdaletechnicalservices.co.uk Data exchange method: Balance indication reading (corresponds to key use): Computer Balance: S I CR LF (53h 49h 0Dh 0Ah) – initiation signal. Balance Computer: balance sends 16 bytes of data according to description: Byte - sign “-” or space Byte - space Byte...
info@tursdaletec chnicalservices.co o.uk Exam ple: To se et 1000g in n balance B1.5 (d=0 .5g) enter: S L 1 1 0 0 0 . 0 0 CR LF (5 53h 4Ch 31 1h 30h 30 h 30h 2Eh h 30h 0Dh 0Ah).
Page 13
info@tursdaletechnicalservices.co.uk Other functions may be available as options for demand (description of all user functions is presented in the separate document). After pressing Menu button, start menu will be displayed. The functions are displayed with the successive numbers: F1-PCS, F2-AUt, ect. User may change the menu content by activating or deactivating the available special functions with the Menu update function (ACt).
info@tursdaletechnicalservices.co.uk 13. Common weighing The 0 key operates only with unloaded pan, and zeroes the balance indication. Weighing results should be read with the " " indicator on. 14. Weighing with tare The balance enables tare setting in the whole measuring range. The B/G button allows for reading gross weight.
info@tursdaletechnicalservices.co.uk 15. Balance resolution increasing The HR button causes the momentary result displaying (approx. secs) with maximum resolution, allowed by the balance processor. That button is especially useful in the legalized balances with d=e reading graduation. The increased resolution result is the auxiliary information, and may not be printed or sent to the computer with the...
info@tursdaletechnicalservices.co.uk 17. Pieces counting function (F..-PCS) That function allows for counting the identical details, e.g. screws or nails, contained in the weighed batch. The measurement takes place in two phases: - first phase – calculation single detail weight upon the basis of sample specified number of pieces 5, 10, 20, 50,...
info@tursdaletechnicalservices.co.uk 18. Setting printer cooperation mode function (F..-LPt) This function sets the balance interface for the cooperation with printer. After its activation the measurement number and result is printed automatically after putting and removing the sample, without pressing key. To enter the computer cooperation mode activation results transmission...
info@tursdaletechnicalservices.co.uk 19. Serial port parameters setting function (F..-rS) Function is available in balances equipped with serial port, e.g. RS232C, and allows for setting the following serial port transmission parameters: - baud rate (bAud: 1200, 4800, 9600), - bits in byte number (bit: 7, 8), - parity control (PArItY: 0, 1;...
info@tursdaletechnicalservices.co.uk 20. Tare entering function (F..-tAr) This function allows for measurement of product gross weight in the container with known weight, and reading of calculated net weight. To do this, first enter the tare value to the balance memory. Entered tare value may be recalled by pressing 0 key, when balance is unloaded.
info@tursdaletechnicalservices.co.uk Measurement with entered tare recalling: During measurement with activated function causes balance zeroing with simultaneous recalling of tare value from balance memory. This value is displayed with sign “-“. allows switching gross weight indications. Repressing the B/G key switches to net weight display Caution: key does not work with unloaded...
info@tursdaletechnicalservices.co.uk 22. Maintenance and small repairs 1. Keep the balance clean. 2. Take care, for during the operation some impurities may enter under balance base and into sensors vicinity. In case of impurities finding remove them. 3. In case of improper operation due to short power failure, switch off the balance by unplugging the power cable from the socket, and plug it again after a few seconds.
info@tursdaletechnicalservices.co.uk Declaration of Conformity PCE-GROUP Europe Im Langel 4 Tel: 0049-[0]2903- 976 99-0 D – 59872 Meschede Fax: 0049-[0]2903-976 99-29 E-Mail: info@warensortiment.de Internet: www.pce-group-europe.com Konformitätserklärung Declaration of conformity for apparatus with CE mark Konformitätserklärung für Geräte mit CE-Zeichen Déclaration de conformité pour appareils portant la marque CE Declaración de conformidad para aparatos con disitintivo CE Dichiarazione di coformitá...