6.2 Installation of the
Transmitted Light Illumination Carrier (TL)
Wipe the installation surface on the microscope
(8.3) with a dry cloth. Tip the illumination carrier
(8.1) back slightly and install it so that the pin (8.2)
engages the groove in the support surface (8.4).
Set the TL illumination carrier upright and fasten
it with the 4 screws.
When fastening the transmitted light illumination
carrier, do not hold it. This will ensure its optimal
alignment with the optical axis.
The tilt angle of the illumination carrier can be
varied with the knurled screw (9.1) or fi xed verti-
cally.
Fig. 8
Installing the transmitted light illumination carrier
1
Transmitted light illumination carrier
2
Transmitted light illumination carrier pin
3
Support surface
4
Support surface groove
5
Support surface groove
6
EXT1-EXT4 sockets
7
Connector cable
1
3
7
2
6
4
Leica DMI4000 B and Leica DMI6000 B
Connect the electronics cable to one of the sock-
ets, EXT1 – EXT4.
The transmitted light lamp housing for 12 V
100 W halogen lamps is a separate component.
For instructions on replacing the halogen lamp
→ Ch. 6.10, p. 45.
Fig. 9
Transmitted light illumination carrier, rear side
1
Knurled locking knob
of the transmitted light illumination carrier
2
Connector cable for the microscope rear side
1
5
6. Assembly
1
2
1
5
31