Problem
No
image
screen) but the cooler
functions
and
digital
image
coherent
No Live image on Altair
No image in external
trigger mode
Altair Image presents
lot of noise
DCO19U-F Titanium User Manual
(black
To check that the boot loader is displayed at
starting on the analog video.
the
correct?
No The monitor (or its connector) is at fault
is
Yes Activate the AGC under CIRRUS. Check the
AGC parameters (Withdraw all the limitations
in gain, offset and ROI). Does video monitor
remain black?
No The problem is solved.
Yes Contact FLIR Systems After Sales Service
Is the camera correctly connected and powered?
No remake connections
Yes Does Cirrus communicate with the camera?
No Contact FLIR Systems After Sales Service
Yes Check that external trigger is deactivated.
Is the problem solved?
No Contact FLIR Systems After Sales Service
Check that synchro signal connected to the
camera is LVTTL (0 – 3.3V) under 50 Ohms.
No Adjust the input signal
Yes Is the frequency of the signal in the range
of the camera (from 3Hz to MAX frequency
proposed by CIRRUS)
No Adjust the frequency of the signal
Yes Replace the external signal by a signal
given by a function generator setup to square
signal from 0 to 3.3V under 50Ohms and
25Hz. Do you get an image ?
No Contact FLIR Systems After Sales Service
Yes Check more precisely your synchronization
signal. Eg. : 50 HZ noise, low voltage under
50 Ohms, Rising time to long, etc.
Is your scene adapted to the temperature range
in use? (eg. : wall at 20°C with a range of 100 -
350°C)
No Adjust the temperature range.
Yes Is the optical path clean ? (lens, filter,
elements between the scene and the camera)
No Be careful that your measurement will not
be coherent.
Yes Contact FLIR Systems After Sales Service
TITANIUM User Manual
Action
Are connections
Page 22