About The Meter; Calibration Values When Using A Millivolt Souce - Omega DP116 User Manual

Thermocouple miniature panel thermometer
Table of Contents

Advertisement

CALIBRATION VALUES
WHEN USING A MILLIVOLT SOURCE
MILLIVOLT
MODEL
VALUE FROM
OPTION
MILLIVOLT
SOURCE
*-JF1
0.0
41.013
*-JF2
0.0
3.917
*-JC1
0.0
41.013
*-JC2
0.0
9.113
*-KF1
0.0
40.449
*-KF2
0.0
2.896
*-KC1
0.0
53.611
*-KC2
0.0
7.737
† Look at the end of Table 4-2, for the explanation of †.
* Look at Table 1-1, for complete models (preface section).
TABLE 4-2
METER
POT TO
DISPLAY
ADJUST
32°F
Zero
1346°F
Span
31.6°F †
Zero
167.0°F
Span
0°C
Zero
730°C
Span
-0.2°C †
Zero
170.0°C
Span
31°F †
Zero
1794°F
Span
32.0°F
Zero
160.0°F
Span
0°C
Zero
1335°C
Span
0.0°C
Zero
190.0°C
Span
26
S S E E C C T T I I O O N N 2 2 A A B B O O U U T T T T H H E E M M E E T T E E R R
2 2 . . 1 1 F F r r o o n n t t o o f f t t h h e e M M e e t t e e r r
Figure 2-1 shows the panel thermometer.
Figure 2-1. Panel Thermometer
Features:
Display: 3 1/2 Digit, 7-Segment Red or Green LED
Full-size 14.2 mm (0.56") LED Display
Analog Output Standard
3/64 DIN Standard Panel Cutout
Removable Screw-Clamp Cable Connector
Display Hold Capability
3