12.4 Error messages

Error message
ERR4
ERR5
ERR6
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
32
Description
Zeroing range exceeded
due to switching-on balance
0
or pressing
(normally
4% max)
Keyboard error
Value outside the A/D
changer range
Possible causes
• Object on the weighing
plate
• Overload when zeroing
• Improper adjustment
• Damaged weighing cell
• Damaged electronics
• Improper operation of the
balance
• Damaged weighing cell
• Damaged electronics
GAB-N-BA-e-1113