Technical Description

General
The TeleMyo 2400T is an electromyographic (EMG) instrument available as 4,8,12, 16, 20 or 24 channel
systems. This device has been manufactured specifically for sEMG applications and is not intended for use
with evoked response equipment. Its design complies with international standard IEC 60601-1, The
International Standard for Medical Electrical Equipment and IEC60601-2-40, Particular requirements for
the safety of electromyographs and evoked response equipment. This device also complies with part 15 of
the Federal Communication Commission (FCC) Declaration of Conformity Statement.
operation is subject to the following two conditions:
1. The device may not cause harmful interference.
2. The device must accept any interference received, including interference that may cause undesired
operation.
Equipment Installation and Restricted Use
The instrument can be charged from all conventional alternating current power sources (i.e. 100-240 VAC
50/60Hz) without modification. A power cord matched to the local service configuration is needed and is
included with the unit. The power cord requires an IEC320 style connector for the instrument end. The
opposite end of the power cord must terminate with a grounded AC plug. Electrically, the TeleMyo 2400T
battery charger is a Class I device, the transmitter is internally powered with type-B applied parts. The
instrument is not protected against the ingress of water and carries no IPX rating (i.e. ordinary equipment).
Restricted Usage
The TeleMyo 2400T is not intended for use in the following situations:
1. In the presence of flammable anesthetizing agents or substances.
2. In the presence of strong magnetic fields, such as those produced by MRI (Magnetic Resonance
Imaging) equipment.
3. In the presence of heavy electrical machinery such as large motors, elevators, compressors, and the
like.
4. Within one-meter proximity of critical medical devices.
5. When the antenna is very close or touching any exposed parts of the body, especially the face or eyes,
while transmitting.
6. With evoked response equipment or high frequency surgical equipment. Connection of a patient to
high frequency surgical equipment and to an electromyograph simultaneously may result in burns at
the site of the TeleMyo 2400T electrode and possible damage to the biological amplifiers.

Transport and Storage

The TeleMyo 2400T is capable of transport and storage in environmental conditions within the following
ranges:
Ambient Temperature: -40 degrees C to +70 degrees C
Relative Humidity: 10% to 100%
Atmospheric Pressure: 500hPa to 1060hPa
P-2308TV2 Rev G (06-06-11)
TeleMyo 2400T V2