Emg Active Lead Connectors 1,3,5 And 7; Right Side; Sync Output Connector; Emg Patient Lead Connectors 2,4,6 And 8 - Noraxon TeleMyo 2400T V2 User Manual

Transmitter/ pc card
Table of Contents

Advertisement

Noraxon U.S.A., Inc.

2. EMG Active Lead Connectors 1,3,5 and 7

Ports to which Active Leads or other qualified analog input signals are connected. These ports come
supplied with protective rubber caps that must be removed to attach inputs cables. The cap must be re-
applied whenever the transmitter is used without an input connection to that channel.

1. Sync Output Connector

The opposite end of the transmitter-to-transmitter cable is connected to this connector. Please refer to
page 4 for instructions.

2. EMG Patient Lead Connectors 2,4,6 and 8

Ports to which Actives Leads or other qualified analog input signals are connected.

3. External Battery Input

An external power source may be used to power the transmitter. It must be 4 to 6 volts with a current
of 400 mA.
Always turn the instrument off (place switch in '0' position) before
connecting the Active EMG Leads. Failure to do so may damage internal
circuitry.
The protective input connector cap must be applied whenever the
transmitter is used without an input connection to that channel. This is
necessary to avoid a potential electrical hazard.
P-2308TV2 Rev G (06-06-11)

Right Side

2
1
TeleMyo 2400T V2
3
29