Charging the battery

To charge the TeleMyo 2400T transmitter's internal battery:
1. Turn the power switch on the transmitter to the '0' or charge position.
2. Unsnap all active leads attached to patient.
3. Remove the transmitter from the patient environment.
4. Plug the battery charger power source into a live wall outlet.
5. Insert the female end into the Battery Charger connector, located on the front display of the transmitter.
While the device begins charging, the Charge indicator may flash if the battery charge is nearly completely
exhausted. As charging continues, the indicator will glow steady. When fully charged the indicator will
turn off. Leave the transmitter plugged in for approximately 120 minutes to completely charge the battery.
When fully charged the battery will run for 4 hours.
The transmitter batteries should undergo a minimum of 300 recharge cycles before needing to be replaced.
Only use replacement battery packs available from Noraxon. Use of unauthorized battery packs may create
a safety hazard.
The transmitter cannot operate while it is recharging. To avoid a
possible electrical hazard, do not have leads attached to a patient while it
is being charged.
The TeleMyo 2400T transmitter must be recharged using the supplied
Transmitter Battery Charger Power Source only. Use of an
unauthorized power source may create a safety hazard.
The opposite end of the power cord has a safety or ground terminal on
its plug. It is essential to only plug the power cord into a wall outlet that
has a properly grounded receptacle at all times to ensure safe operation
and full compliance with the relevant standards.
P-2308TV2 Rev G (06-06-11)