INTRAORAL IMAGING PROCEDURE
IMPORTANT:
Scanning and Erasing Intraoral Plates. Scan and erase an intraoral PSP in one operation as follows .
1 . Carefully open the Plate Transfer Box .
2 . Grasp a PSP by the edges between your thumb and index finger .
3 . With the printed side facing you, carefully and quickly insert the PSP into the corresponding plate guide
slot as far as possible . The wide side of Size #2 plates are fed into the guide as shown by Figure 12 .
Plate sizes #0, #1 and #3 are fed with the short side into the guide .
4 . Immediately press the PSP all the way down until the scanner's transport mechanism takes over and
the PSP moves down on its own .
5 . Verify that both the corresponding lane and pointer LED indicators light amber, indicating the Plate
Guide slot is in use processing the plate .
PSP Pick-up.
Note: Up to four PSPs can be processed simultaneously. One PSP can be inserted
into each Plate Guide at a time as long as the corresponding track indicator
light is illuminated bright green. The next PSP may be fed into a Plate Guide slot
only after the corresponding track indicator LED changes from amber to green.
6 . Observe that a red glow emanates from the scanner exit slot .
7 . Repeat steps 2 through 6 to process additional PSPs as necessary . Another PSP may be fed into
any Plate Guide slot as long as the track indicator light illuminates green .
8 . Observe that each scanned and erased PSP drops into the receiving tray at the bottom of the
scanner . Since the default operation is with the erase mode enabled (ERASER Status LED Indicator
lit steadily blue), each PSP is erased and ready for reuse .
Note: The lane and pointer LED indicators remain lit green as long as the imaging
software is set to acquire images.
9 . Observe that the associated track indicator light illuminates amber when processing PSPs . Verify
that the track indicators return to green when the processing ends .
10 . Also verify that the red glow from the exit slot extinguishes after the last PSP drops to the tray .
CAUTION: Always use exercise care in handling the PSPs so as not to scratch the
sensitive surface or nick the edges.
11 . Retrieve the processed (scanned and erased) PSPs for reuse or storage . Make sure not to scratch
the sensitive surface or nick the edges when removing from the scanner outlet .
12 . View and save each scanned image using the user-supplied imaging software .
Page 20
Transfer the PSP from the Plate Transfer Box to the Plate Guide slot quickly. Always keep
the sensitive side away from any light source to minimize image loss.
Feeding Intraoral Plate.
Figure 12 . Feeding an Intraoral Plate
Fully Inserted Plate.