Waste Heat/Required Cooling Performance; Features; Overview Of Usable Vis/Uv Lasers - Leica TCS SP5 User Manual

Hide thumbs Also See for TCS SP5:
Table of Contents

Advertisement

2.7.5

Waste Heat/Required Cooling Performance

The TCS SP5 system has a maximum power consumption of 3.2 kW (VIS system) or 6.2 kW
(MP system).
The TCS SP5 X has a maximum power consumption of 3.4 kW.
For the specifications of external lasers such as UV and MP lasers, please refer to the
manufacturer's documentation.
2.8

Features

2.8.1

Overview of Usable VIS/UV Lasers

The Leica TCS SP5 features a combination of the lasers listed below.
Wavelength
Laser type
Diode 405
405
Diode 405 p
405
DPSS 445
445
458, 476, 488,
Ar
496, 514
458, 476, 488,
Ar, external
496, 514
HeNe
543
DPSS 561
561
HeNe
594
HeNe
633
UV, external
355
Table 1: Usable lasers for TCS SP5 (without MP)
Maximum
luminous power
at laser output
[nm]
[mW]
< 120
< 5 (mean
power)
< 75
< 200
< 500
< 1.5
< 100
< 4
< 15
< 500
Maximum
luminous power
in focal plane
[mW]
Continuous wave
< 7
(cw)
< 0.3 (mean
pulsed, 60 ps
power)
Continuous wave
< 7
(cw)
Continuous wave
< 50
(cw)
Continuous wave
< 125
(cw)
Continuous wave
< 0.5
(cw)
Continuous wave
< 12
(cw)
Continuous wave
< 1
(cw)
Continuous wave
< 5
(cw)
Continuous wave
< 18
(cw)
Pulse duration
19