Error Messages

Your meter will explain what an error message means, its possible causes,
OK
and what to do about it.
There is a problem
Notes:
with the meter.
Remember to tap the arrow button to see all of the important
Call customer service
information.
for help and reference
Error 1, or get more
Tap OK to return to the Home screen.
information below.
1 of 5
Example of an Error Message
Temperature Warning
This symbol may appear after inserting a test strip. It will include a
message saying the meter is either too hot or too cold to use right now.
Blood glucose tests taken outside of the operating temperature range
may not be accurate.
Tap OK to clear the screen and test again later.
17
Control Solution Test
A control solution test should be performed when you are not sure of your results and want to
confirm that your meter and test strips are working correctly.
IMPORTANT:
Control solution results should fall within the control solution range printed on the test
strip vial.
Do not use control solution past its expiry date. Discard control solution 3 months after
opening or on the expiry date printed on the bottle, whichever comes first.
(Example: open April 15, discard July 15; write the discard date on the side of the bottle.)
The control solution range is a target range for control solution only, not for your blood
glucose level.
Use only FreeStyle control solution with this meter.
Replace the cap securely on the bottle immediately after use.
Do not add water or other liquid to control solution.
There are two levels of control solution available (Low and High).
Contact Customer Service for information on how to obtain control solution.