3. During charging the battery capacity display shows the progress of the refresh
procedure in the battery. The time required will vary according to the capacity
of the battery (9-12 hours).
D. IF A PROBLEM ARISES DURING CHARGING
If a problem occurs in the battery or in the charger during charging, this is indicated
by the alternate flashing charge displays (green) and (yellow) on the Refresh display
on the charger and/or on the battery capacity display.
The displays on the charger and the charger Refresh display flash simultaneously.
This indicates an internal malfunction in the battery. Please replace the battery with a
new one.
The fourth LED on the battery capacity display flashes.
This indicates an internal malfunction in the battery. Please replace it with a new one.
The charge display and the refresh display on the charger flash rapidly (in 1-sec-
ond intervals), and the third LED on the battery capacity display flashes.
• The contacts may be malfunctioning. Please
check if small objects have broken the electrical
contact between the battery terminals and
those of the charger, and remove them. If there
is dirt or dust on the contacts, remove it with a
dry cloth, ensuring that the contacts are not
bent or damaged
• The battery fuse may have blown. Open the fuse
cover (A) on the side of the battery and check the
fuse. If the fuse is damaged, fit the replacement
fuse which is located on the protective cap.
There is a fire risk if the authorized fuse (20 A fuse) is not used. Do not use
any other. Never use wire instead of the authorized fuse.
The first, third and fifth LED on the battery capacity display will flash.
The battery is not at the correct temperature for charging. Wait until the battery is at
a suitable temperature and begin the charging process.
The charging lights and the refresh lights flash alternately in a slow sequence (2-
second intervals).
The charger is faulty. Replace it with a new one and carry out the charging process.