G. REDUCTION OF BATTERY CAPACITY
If you press the display key and not all the LEDs on the battery capacity display are
illuminated after charging, this shows that the battery capacity is reduced.
Depending on the condition of the battery, the capacity display can return to maxi-
mum with successive charging processes.
Battery status:
Battery capacity is reduced to 60 to 80%. Sufficient for normal use of power drive.
Verdict:
The battery capacity is reduced in comparison to that of a new one. Replacement with a
new battery is necessary, if you want to travel a long way on only one battery charge.
Battery status:
The battery is capable of storing at most 60% of its original capacity. Adequate use of
the power drive is no longer possible.
Verdict:
Replace the battery with a new one. Time for replacement.
NOTE– • Depending on the condition of the battery, the capacity display can return to
the original maximum capacity following successive charges.
• Contact your dealer when no more than 3 LEDs on the capacity display light
up after the charging process.
H. BATTERY MALFUNCTION DISPLAY
If there is an internal malfunction in the battery or if the battery temperature lies
outside the acceptable range, information regarding battery status will appear on the
battery capacity display.
Internal battery malfunction
• The fourth LED flashes: please contact your dealer.
• Although charging cannot be carried out, it is possible to use the battery until
it is completely discharged.