25
NO PRESS IN HYD SYS: FOAM
26
EVF LEAKAGE
27
EVC LEAKAGE
28
EVD LEAKAGE
29
WASHING
FAILURE
30
SAFETY CHAMBER LEVEL EXCEEDED
31
SAFETY LEVEL FAILED
32
SUMP FULL
33
LACK
CONDENSER
34
CONDENSER DRAINAGE FAILED
35
CONDENSER LEVEL SWITCH FAILED
36
CONDENSER DRAINING PUMP FAILED
37
NOT REACHED TARGET DRAINAGE AT
T=
38
COOLED DRAINAGE FAILED
41
DETERGENT 1 INFLOW FAILED
(flowmeter nr.1 optional)
42
DETERGENT 2 INFLOW FAILED
(flowmeter nr.2 optional)
43
DETERGENT 3 INFLOW FAILED
(flowmeter nr.3 optional)
44
DETERGENT 4 INFLOW FAILED
Versione 22.10.08
Manual GW4090 - User
CHAMBER
OF
WATER
IN
Check the type of detergent used.
Repeat the cycle.
Check the water supply (cock
open/closed, water pressure,
connection pipe, etc.)
Repeat the cycle. Call the Technical
Assistance Service if the fault persists.
Check the water supply (cock
open/closed, water pressure,
connection pipe, etc.)
Repeat the cycle. Call the Technical
Assistance Service if the fault persists.
Check the water supply (cock
open/closed, water pressure,
connection pipe, etc.)
Repeat the cycle. Call the Technical
Assistance Service if the fault persists.
Make sure that the drain pipe is
DRAINAGE
positioned as indicated in the manual.
Repeat the cycle. Call the Technical
Assistance Service if the fault persists.
Repeat the cycle. Call the Technical
Assistance Service if the fault persists.
Make a Reset cycle. Repeat the cycle.
Call the Technical Assistance Service
if the fault persists.
Check the water supply (cock
THE
STEAM
open/closed, water pressure,
connection pipe, etc.)
Repeat the cycle. Call the Technical
Assistance Service if the fault persists.
Make sure that the drain pipe is
positioned as indicated in the manual.
Repeat the cycle. Call the Technical
Assistance Service if the fault persists.
Repeat the cycle. Call the Technical
Assistance Service if the fault persists.
Repeat the cycle. Call the Technical
Assistance Service if the fault persists.
Repeat the cycle. Call the Technical
Assistance Service if the fault persists.
Make sure that the suction tubes of can
P1 are in a good condition and not
squashed. Repeat the cycle. Call the
Technical Assistance Service if the
fault persists.
Make sure that the suction tubes of can
P2 are in a good condition and not
squashed. Repeat the cycle. Call the
Technical Assistance Service if the
fault persists.
Make sure that the suction tubes of can
P3 are in a good condition and not
squashed. Repeat the cycle. Call the
Technical Assistance Service if the
fault persists.
Make sure that the suction tubes of can
P4 are in a good condition and not
39