Symptoms of High or Low Blood Glucose

You can better understand your test results by being aware
of the symptoms of high or low blood glucose. According to
the Canadian Diabetes Association (www.diabetes.ca), some
of the most common symptoms are:
Low blood glucose
(Hypoglycemia):
• shakiness
• sweating
• fast heartbeat
• blurred vision
• confusion
• passing out
• seizure
• irritability
• extreme hunger
• dizziness
Ketones (Ketoacidosis):
• shortness of breath
• nausea or vomiting
• very dry mouth
WARNING
If you are experiencing any of these symptoms,
test your blood glucose. If your test result is under
2.8 mmol/L or above 13.9 mmol/L, contact your
health care professional immediately.
For additional information and a complete list of symptoms,
contact your health care professional.
®
62
High blood glucose
(Hyperglycemia):
• frequent urination
• excessive thirst
• blurred vision
• increased fatigue
• hunger

Specifications

Test Sample: Venous or capillary whole blood
Test Result: Referenced to plasma/serum glucose
Sample Volume: 0.6 μL
Measuring Range: 0.6 - 33.3 mmol/L
Countdown Time: 5 seconds
Memory: Stores most recent 800 test results
Battery Type: Two 3-volt lithium batteries
(DL2032 or CR2032)
Battery Life: Approximately 1000 tests (1 yr. average use)
Operating Temperature Range: 5°C - 45°C
Humidity: 10–93% RH
Dimensions: 80 mm (H) x 53 mm (W) x 12.7 mm (T)
Weight: 1.6 oz. (45 grams)
Meter Life: 5 years
Sound Output: 55 to 80 dBA at a distance of 10 cm
Electromagnetic Compatibility (EMC): The C
meter complies with the electromagnetic requirements
specified in ISO 15197: 2013. Electromagnetic emissions
are low and unlikely to interfere with other nearby electronic
equipment, nor are emissions from nearby electronic
equipment likely to interfere with the C
The C
N
meter meets the requirements of
ONTOUR
EXT
IEC 61326-2-6 for immunity to electrostatic discharge. It
is advisable to avoid use of electronic devices in very dry
environments especially if synthetic materials are present.
The C
N
meter meets the requirements of
ONTOUR
EXT
IEC 61326-1 for radio frequency interference. To avoid
radio frequency interference do not use the C
meter near cellular or cordless telephones, walkie talkies,
garage door openers, radio transmitters or other electrical
or electronic equipment that are sources of electromagnetic
radiation, as these may interfere with the proper operation of
the meter.
®
N
ONTOUR
EXT
N
.
ONTOUR
EXT
N
ONTOUR
EXT
63