13.3 Calibration of the balance using test weights

Exercise particular care when carrying out the calibration procedure.
The GEO values are overwritten following calibration using test weights.
These may then no longer be used.
ON
OFF
2s
0
F
F
F
52
when
"test"
appears
PASS
0 - - -
Dotted segments flashing
Selection of the service
parameter menu.
Selection of balance
parameters.
Hidden entry of service pass-
word.
Programme start is
carried out if a pass word is
not entered.
Pass code is 0001. The
positions are selected using
0
.
The figures are entered using
Adopt and check service
menu.
Balance completely installed
with load plate/pre-load and
stabilized
0 % max.
Zero calibration,
Zero calibration until the count
of the counter no longer
changes.
Apply max. load to the
balance
100 % max.
Range calibration,
Calibration of range until the
count of the counter no
longer changes.
F
Use
to
close/confirm