14.2 Cleaning

Pressure compensation
The weighing terminal has a pressure compensation device for the weighing cell.
This is located next to the measuring cable inlet and consists of a fixed connecting
piece with adhesive diaphragm and a screw plug.
There are three openings on the circumference of the fixed connecting piece and
these must not be clogged by dirt.
Depending on the degree of contamination, these openings must be cleaned at
certain intervals.
Unscrew the hexagonal screw plug and remove any dirt.
Be particularly careful not to damage the diaphragm.
Hand-tighten the screw plug after cleaning.

14.2.1 Weighing terminal

Clean the weighing terminal from time to time using a damp cloth.
Do not use aggressive cleaning agents.
Provide the unit with a water-tight cover when cleaning by means of a
pressure washer.