Do not overfill the humidifier as water may enter the device and air tubing.
Leave the humidifier to cool for ten minutes before handling to allow the water to cool and
to make sure that the humidifier is not too hot to touch.
Make sure that the humidifier is empty before transporting the device.
Technical specifications
Units are expressed in cm H
90W power supply unit
AC input range:
DC output:
Typical power consumption:
Peak power consumption:
Environmental conditions
Operating temperature:
Operating humidity:
Operating altitude:
Storage and transport temperature:
Storage and transport humidity:
Electromagnetic compatibility
The AirCurve 10 complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-
2:2007, for residential, commercial and light industry environments. It is recommended that mobile communication devices
are kept at least 1 m away from the device.
Information regarding the electromagnetic emissions and immunity of this ResMed device can be found on
www.resmed.com, on the Products page under Service and Support.
EN 60601-1:2006 classification
Class II (double insulation), Type BF, Ingress protection IP22.
Sensors
Pressure sensor:
Flow sensor:
Maximum single fault steady pressure
Device will shut down in the presence of a single fault if the steady state pressure exceeds:
30 cm H
O (30 hPa) for more than 6 sec or 40 cm H
2
Sound
Pressure level measured according to EN ISO 17510-1:2009 (CPAP mode):
SlimLine:
Standard:
Power level measured according to EN ISO 17510-1:2009 (CPAP mode):
SlimLine:
Standard:
Declared dual-number noise emission values in accordance with ISO 4871:1996.
32
O and hPa. 1 cm H
2
2
O (40 hPa) for more than 1 sec.
2
O is equal to 0.98 hPa.
100–240V, 50–60Hz 1.0–1.5A, Class II
115V, 400Hz 1.5A, Class II (nominal for aircraft use)
24V
3.75A
53W (57VA)
104W (108VA)
+5°C to +35°C
Note: The air flow for breathing produced by this therapy
device can be higher than the temperature of the room.
Under extreme ambient temperature conditions (40°C) the
device remains safe.
10 to 95% relative humidity, non-condensing
Sea level to 2,591 m; air pressure range 1013 hPa to
738 hPa
-20°C to +60°C
5 to 95% relative humidity, non-condensing
Internally located at device outlet, analogueue gauge
pressure type, -5 to +45 cm H
Internally located at device inlet, digital mass flow type, -70
to +180 L/min
26.6 dBA with uncertainty of 2 dBA
26.6 dBA with uncertainty of 2 dBA
34.6 dBA with uncertainty of 2 dBA
34.6 dBA with uncertainty of 2 dBA
O (-5 to +45 hPa)
2