Reprocessing the air tubing
Disconnecting
1. Hold the cuff of the air tubing and gently pull it away from the device.
2. Hold both the cuff of the air tubing and the swivel of the mask, then gently pull apart.
Decontaminating
Before the disinfection process, each component must be cleaned and rinsed so no visible
contamination is present.
1. Clean all components with a soft bristled brush for one minute while soaking in detergent
solution (see table below). Pay particular attention to all crevices and cavities.
2. Run the detergent solution through the air tubing repeatedly until no contamination is visible.
3. Thoroughly rinse each component according to the detergent manufacturer's instructions.
ResMed has tested the following detergents according to the manufacturer's instructions:
Detergent
™
Alconox
(diluted at 1%)
™
Neodisher MediZym
(diluted
at 2.0%)
Disinfecting
In the procedures below, only one disinfection process needs to be performed.
High level thermal
Part
SlimLine
ClimateLineAir
Standard
1. Immerse the air tubing in a water bath.
Take care that no air bubbles are trapped inside the air tubing.
2. Increase the water bath temperature to 70°C for 100 minutes, or a maximum of 75ºC for
30 minutes. Higher temperatures may damage the tubing.
3. Air dry out of direct sunlight and/or heat.
Water temperature
Hot water (approx 60°C)
Warm water (approx 45 to 60°C)
Room temperature water (approx 21°C)
Warm water (approx 45 to 60°C)
Validated number of cycles
Hot water: 75°C for 30 minutes OR 70°C for 100 minutes.
20
10
100
SlimLine ClimateLineAir Standard
English
19