INTRODUCTION RESLINK'" a device designed for use with ResMed flow generators to help your clinician monitor your sleep and ensure that you are receiving optimal treatment. When connected flow generator, RESLINK records treatment and pulse oximetry data during therapy. The data is stored on a SmartMedia card-a memory card capable of storing up to ™...
Page 8
Do not att e mpt to dismantle the RESLINK. There are no user viceable parts • inside. Repairs and internal vicing should only be performed by an authorised agent There a serial (9-pin) port the back of some flow generators.
RES LINK Ensure that flow generator is powered off before attaching the RESLINK. Use the switch the back of the flow generator to power off the device. 2. Slide the lever at the...
Only the • Sensors are compatible with the Reslink. Connecting other oximeters may cause personal injury or damage to the flow generator or the Reslink. Check the oximeter cables and the oximeter sensor before use. If there is any •...
LCD. When the following message displayed, you know that the Reslink is correctly connected. displayed will The number reflect the Reslink software version 2. If necessary. the software in your flow...
If attached. remove the oximeter sensor f rom your finger. the last After stopping therapy night that you use the RESLINK, wait minutes then tum off your flow generator at the power switch.
These messages will be interspersed with other flow generator messages. Successful Reslink connection Indicates that FOUND CARD SmartMedia Card present ResLink 1 /4 that...
Page 14
These messages will be interspersed with other flow generator messages. Successful connection Reslink Sp02: 98% HR: 65 First oximetry message. OXIMETER are 6 There message s Each...
• off . If an oximeter connected, pull the oximeter plug out of the RESLINK oximeter port (1-1). RES LINK flow generator power off using the power switch. (E-2). Slide the lever at the...
CLEANING AND MAINTENANCE CAUTION Avoid exposing the Reslink to moisture and extreme temperatures. CLEANING THE RESLINK necessary, wipe outside of the RES LINK with a clean cloth. CLEANING THE OXIMETER . If necessary, wipe surfaces with a soft cloth dampened...
Do attempt to open the RESLINK. Problem Possible Cause Solution Difficulty attaching the Incompatible flow The RESLINK AND 58 RESLINK are RESLINK to flow generator. compatible for use with certain generator. ResMed flow generators only. "Compatible flow generators"...
Page 18
For the S8 RESLINK make sure that the top and bottom latches are secured properly. Incompatible flow The RESLINK and S8 RESLINK are generator. compatible for use with certain ResMed flow generators only. See "Compatible flow generators" page Faulty RESLINK.
Page 19
Stop therapy. Power the flow off . UPGRADE FAILED delivering therapy. generator Check the power supply to the flow generator. Power the flow generator back "Using the Reslink" on page Difficulty inserting the Incompatible oximeter Ensure that a NONIN (Model oximeter into the RESLINK.
Page 20
ResLink should then start normal SmartMedia Card. operation. The following message ResLink has been removed Attach the Reslink or 58 ResLink displayed on the LCD: from flow generator to the flow generator. Make sure not attached correctly. that it is properly attached.
Dimensions (H x W X 20mm x 65mm 65mm Weight: 14 g Power Supply: Maximum power consumption derived from the flow generator. 400mA RESLINK AND 58 RESLINK Housing Construction: Flame retardant engineering thermoplastic. IEC 60601-1 Class (double insulation). Type CF Classification: Electromagnetic Compatibility...
Page 22
ResMed device please contact your ResMed office, local distributor g o to www.resmed.com/environment. Not all flow generators are available regions...