Supplemental oxygen
The AirCurve 10 device is designed to be compatible with up to 15 L/min of supplemental oxygen in
CPAP and ASV modes and up to 4 L/min in ASVAuto mode.
At a fixed rate of supplemental oxygen flow, the inhaled oxygen concentration will vary depending
on the pressure settings, patient breathing pattern, mask selection and the leak rate.
To connect supplemental oxygen to the device you need to connect an oxygen connector port. For
more information on how to set up the device with supplemental oxygen, refer to the user guide
supplied with that accessory.
Notes:
Adding oxygen may affect the delivered pressure and the accuracy of the displayed leak and
minute ventilation.
Before adding oxygen, familiarise yourself and your patient with the specific warnings relating to
the use of supplemental oxygen. These can be found at the end of this guide.
Antibacterial filters
Antibacterial filters increase resistance in the air circuit and may affect accuracy of displayed and
delivered pressure, particularly at high flows.
ResMed recommends using an antibacterial filter with a low impedance [eg, 2 cm H
60 L/min], such as PALL (BB50T), Air Safety Filter without Luer Port (4222/702) or Air Safety Filter
with Side Port 24966 (4222/701). If using the Air Safety Filter with Side Port, an Oxygen Connector
Port is required.
Note: When using the SlimLine air tubing above 20 cm H
performance may not be reached if used with an antibacterial filter. The device performance must
be checked prior to prescribing the SlimLine air tubing for use with an antibacterial filter.
8
O (20 hPa), the device optimum
2
O (2 hPa) at
2