11. Disassembling And Reassembling Procedures For C·mvr - Olympus CH Series Repair Manual

Ch series
Table of Contents

Advertisement

10·2-8 Adjust the guide
a)
Tighten two Adjusting screws HU3x4SA by
using Torque screwdriver OT0044.
(See Fig. 10-16)
Tightening force: 700 g·cm
b)
Finally tighten three Screws AB3x10SA on
the outer gu ide.
NOTE.
For later steps, proceed to Section 5 of this
manual.
-
- ; .
Fig. 10·16
11. DISASSEMBLING AND REASSEMBLING PROCEDURES FOR C·MVR
J
QJ
Fig.11·1
o
AA877900
AA878000
o
3PSK2x6SA
11-1 Disassembly
procedure
of
East-West
Drive
guide
11·1·1 Remove Specimen holder unit by drawing
4 Screws 3PSK2x6SA.
(See Fig. 11·1)
11-1-2 Remove
Rack
AA877900 by drawing 3
Screws 3PUK2x6SA.
(See Fig. 11-1)
11-1-3 Remove
East·West
Graduated
plate
AA-
878000 by drawing 3 Screws PUK2x3SA.
(See Fig. 11-2)
---------- --------
Fig. 11·2
PUK2x3SA
-45-