8.
REASSEMBLING PROCEDURES FOR C-CL LIGHT EXIT
8-6 Tighten
Ring
AA871600 by using Wrench
KKAA8716.
8-1
Reassemble Lens AA410900 with Lens tube
AA871500.
8·7
Loosen Ring AA871600 to such a slight degree
as not to allow the lens to play, and apply
Araldite OT1028 at a point to prevent loosen-
ing.
(See Fig. 8·1)
8-4
Reassemble Ring AA290100.
8-5
Reassemble Lens LA41 0800.
8-2
Reassemble Ring AA290200.
8·3
Reassemble Filter LP0721 00.
-
AA290200
~
LP072,00
AA290,00
LA41 0800
I
AA87,600
AA87,SOll
_LA410900
9.
OVERALL ASSEMBLY AND ADJUSTMENT
9-1
Reassemble C·CH unit with the arm by using
three
Screws A83x 10SA and three Screws
AA800300.
See step 3-3 above.
9-2 Adjust perpendicularity of the stage .
9-2-2
Adjustment in Y direction
a)
Proceed as described in 9-2-1·b) above.
b)
Proceed as described in 9-2·1-e) above.
(See Fig. 9·2)
c)
If the standard is not satisfied, adjust by
inserting tin foil between Stage plate AA-
871800 and Block AA870500.
Adjust in X direction
Set Stage perpendicularity gauge SKN0003 on
the nose piece.
Apply SKN0003 on the stage surface in the
X direction.
Measure gaps in the X direction by using
Thickness gauge OT0317 between SKN0003
and stage surface.
(See Fig. 9-1)
Standard: Within 0.15 mm
If the standard is not satisfied, adjust position
of C-CH by turning three Setscrews A83x
10SA.
b)
c)
d)
9-2-1
a)
/
Fig. 9-1
Fig. 9-2
- 38-