Iddc Error; No Remote Error; Scanning Program - Keithley 705 Instruction Manual

Table of Contents

Advertisement

1 lddC
I
J
a. Illegal Device Dependent Command (IDDC)
1
1
1
IddCcI
1
1
I
b. Illegal Device Dependent Command Option (IDDCO)
p7-l
c. No Remote
Figure 3-9. IEEE Display Error Message
3.9.1 IDDC Error
An IDDC error results when the Model 705 receives an invalid
command such as $X. This command
is invalid because no such
letter exists in the instrument's
programming
language.
Programming Example-To
demonstrate
an IDDC error, enter
the following statements
into the HP-85 keyboard:
REMOTE 717 (END LINE)
OUTPUT 717; "$X" (END LINE)
When END LINE is pressed the second time, the error message
in Figure 3-Qa is displayed for about one second.
3.9.2 No Remote Error
A front panel no remote error message
will be displayed
if the
Model 705 is not in the remote mode when it receives
a com-
mand over the bus. If an attempt is made to program the instru-
ment when it is not in the remote mode, the no remote message
in Figure 3-9~ will be displayed
on the front panel for about one
second,
Programming Example
-To
make sure the instrument
is not in
the remote mode, enter the following statement into the HP-85:
LOCAL 7 (END LINE)
Now enter the following
programming
statement
into the key-
board:
OUTPUT 717; "Dl X" (END LINE)
When this statement
is executed,
the no remote error message
in Figure 3-9c is displayed on the front panel of the instrument for
about one second.
3.10 SCANNING PROGRAM
The following
program
sets the Model
705 for the following
parameters.
1. Sets programmed
first channel to 1.
2. Sets programmed
last channel to 20.
3. Sets interval time to one second.
4. Sets the unit to the display mode.
5.
Use the continuous
scan mode.
6. Trigger the scan sequence on "X".
7.
Provide a pause and clear statement.
Enter the following
statements
into the HP-85 to program
the
Model 705 to scan chgnels
1 through 20 at a one second inter-
val rate.
PROGRAM
COMMENTS
IO
20
30
35
40
50
60
70
80
REMOTE 717 (END
LINE)
CLEAR 7 (END LINE)
OUTPUT 717;
"DOFl L2OWl P2T4X"
(END LINE)
PAUSE (END LINE)
ENTER 717; A$ (END
LINE)
DISP A$ (END LINE)
PAUSE (END LINE)
CLEAR 7 (END LINE)
END (END LINE)
Sets the Model 705 to the
remote mode.
Clears the Model 705.
Programs the Model 705 for the
above parameters.
Obtain data string.
Display data string.
Program pause.
Clear 705.
End of program.
To start the program press the RUN key on the HP-85. After the
RUN key is pressed the Model 705 scans the programmed
chan-
nels at the programmed
interval rate. To stop the program simply
press the CONT key. The program will stop at the presently
dis-
played channel when the CONT key is pressed. To resume the
program press the CONT key a second time or press the RUN
key. Upon pressing the RUN key or the CONT key (second time)
the Model 705 continues
the program
from the presently
dis-
played channel.'
3-25