Keithley 705 Instruction Manual page 19

Table of Contents

Advertisement

Steps 1 through
7 show
how to get the Model 705 into a
scanning
sequence.
The next examples
show
how to use
the different
scan modes,
how to daisy chain several Model
705's for an extended
number
of channels,
and scanning
using different
scanner
cards,
Example
1 Manual
Scan-The
Model 705 can be program-
med for the step scan mode, if it is desired to scan through
the
programmed
channels
manually.
In the step scan mode the
programmed
channels
can be scanned
(closed
for the pro-
grammed
interval rate and then opened)
one channel at a time.
Upon the actuation
of the START/STOP
button a single chan-
nel will be scanned.
For this example,
choose
the following
parameters:
1. First channel
is 1.
2. Last channel
is 10.
3. Interval
time is two seconds.
4. Number
of poles is two.
5. Use the step scan mode.
Use the following
procedure
to program
the Model 705 for the
preceding
parameters:
1. Install the desired scanner card into the Model 705. Refer to
paragraph
2.2.2.
2. Select the appropriate
line voltage
and turn the Model 705
on. Refer to paragraphs
2.2.1 and 2.3 respectively.
3. Press PRGM, 6, 2, ENTER. For more complete
information
refer to paragraph
2.6.7.
4. Press 0, 1, FIRST. (Programs
channel 1 as the first channel.)
5. Press
1, 0,
LAST.
(Programs
channel
10 as the
last
channel.)
6. Press INTERVAL,
0, 0, 2, 0, 0, 0, ENTER. (Programs
the in-
terval time for two seconds.)
7. Press CHANNEL,
RESET. (Selects
the channel
mode for
display and resets to the first channel.)
8. Press the START/STOP
button
to initiate
the step scan
mode.
Upon
the actuation
of the START/STOP
button
in step 8,
channel
1 is closed and the START/STOP
LED is turned
on.
Channel 1 remains closed for two seconds (programmed
inter-
val rate) and then opens,
The START/STOP
LED turns off
and the Model 705 advances
to the next channel.
To scan the
remaining
channels
press START/STOP
button
each time a
channel is to be scanned.
When the last programmed
channel
is scanned,
in this example,
it goes through
the same pro-
cedure
as previously
stated.
The difference
is that after the
channel
opens up and turns off the START/STOP
LED, the
Model 705 resets to the first programmed
channel.
To stop the scanning
sequence
and reset the Model 705 to the
programmed
first channel
simply
press the RESET button.
Example
2 Single
Scan-In
the Single
Scan mode
all the
channels are scanned one time. The Scan runs as follows:
the
first channel
is closed for the programmed
interval
rate and
then opened.
The Model
705 advances
to the next channel,
closes it for the programmed
interval
rate and then opens the
channel. The sequence
is repeated
up to and including
the pro-
grammed
last channel,
at this point the Model 705 reverts to
the programmed
first channel
and stops. Upon the actuation
of the START/STOP
button
all the programmed
channels are
scanned at the programmed
interval
rate. For this example
let
us choose the following
parameters:
1. First channel
is 3.
2. Last channel
is 15.
3. Interval time is 0.5 seconds.
4. Number
of poles is 2.
5. Use the single scan mode.
Use the following
procedure
to program the Model 705 for the
preceding
parameters:
1. Install the desired scanner card into the Model 705. Refer to
paragraph
2.2.2
2. Select appropriate
line voltage
and turn the Model 705 on.
Refer to paragraphs
2.2.1 and 2.3 respectively.
3. Press PRGM, 6, 2, ENTER. For more complete
details refer
to paragraph
2.6.7.
4. Press 0, 3, FIRST. (Programs
channel 3 as the first channel.)
5. Press 1, 5,
LAST.
(Programs
channel
15 as the
last
channel.)
6. Press INTERVAL,
0, 0, 0, 5, 0, 0, ENTER (Programs
the in-
terval time for 0.5 seconds.)
7. Press CHANNEL,
RESET. (Turns on the CHANNEL
LED,
selects the channel
mode for display and resets to the pro-
grammed
first channel.)
8. Press SINGLE.
(Selects the single scan mode and turns on
the SINGLE LED.)
9. Press the START/STOP
button
to inititate
the single scan
mode.
Upon the actuation
of the START/STOP
button
in step 9,
channel 3 is closed for 0.5 seconds (programmed
interval
rate
and the START/STOP
LED turns on). Channel 3 is opened
and the Model 705 advances
to the next channel.
Channel 4 is
closed for 0.5 seconds
and then opens. The Model 705 then
advances to channel 5 and the cycle repeats itself until the last
channel
is scanned
(closed for the programmed
interval
rate
and then opened).
At the end of the scan of the last channel,
the Model
705 resets to the programmed
first channel
and
turns off the START/STOP
LED. At this point the single mode
has been completed.
Pressing the START/STOP
button
during a single scan stops
the scan at the presently
displayed
channel.
The channel
re-
mains closed and turns off the START/STOP
LED. To start
the scanning
sequence
from the presently
displayed
channel
press the START/STOP
button.
To stop the scanning sequence and reset the Model 705 to the
programmed
first channel simply press the RESET button.
Example
3 Continuous
Scan-In
the Continuous
scan mode
all the programmed
channels
are scanned
continuously.
The
scan runs as follows:
the first channel
is closed for the pro-
grammed
interval
rate and then opened.
The Model 705 then
advances to the next channel,
closes it for the programmed
in-
terval
rate and then
opens
the channel.
The sequence
is
2-8

Advertisement