Keithley 705 Instruction Manual page 21

Table of Contents

Advertisement

3. Select appropriate
line voltage
and turn the Model 705's
on. Refer to paragraphs
2.2.1 and 2.3 respectively.
4. Press PRGM,
6, 2, ENTER. For complete
details refer to
paragraph
2.6.7.
5. Program
one Model
705 as the master by initiating
front
panel
Program
91. For complete
details
concerning
the
front
panel programs
refer to paragraph
2.6.
6. Program
remaining
Model 705's as slaves by initiating
front
panel Program,
92 for each of the remaining
Model 705's.
To do this press PRGM button
and then press 9, 2.
All remaining
Model 705's will have to be programmed
in
the same way.
For complete
details concerning
the front
panel programs
refer to paragraph
2.6.
NOTE
Initiating
front
panel
Program
92 locks
up the
Model 705 front panel except for program 90 and
91 which
allow
you
to go to stand
alone
or
master. The following
message is displayed
while
the front
panel is locked up.
I i;Lf?UE I
I
I
7. Press 0, 1, FIRST. (Programs
channel
1 as the first chan-
nel.)
8. Press 6, 0, LAST.
(Programs
channel 60 as the last chan-
nel.)
NOTE
If the Model 705's are not daisy chained together,
pressing 6, 0 in step 8 causes the Model
705 to
display the following
message for approximately
0.5 seconds.
-1
The display
message
indicates
that there
is no
channel 60.
9. Press INTERVAL,
0, 0, 2, 0, 0, 0, ENTER. (Programs
the
interval
time for two seconds.)
10. Press CHANNEL,
RESET.
(Turns
on the channel
LED,
selects the channel
mode for display and resets the Model
705 to the programmed
first channel.)
11. Press CONTINUOUS.
(Selects the continuous
scan mode
and turns on the continuous
LED.)
12. Press the START/STOP
button
to initiate the continuous
scan mode.
Upon the actuation
of the START/STOP
button
in step 12,
the continuous
scan sequence
starts from
the programmed
first channel.
The continuous
scan mode operates
in the same
way as in Example 3. That is, the first channel is closed for the
programmed
interval time and the START/STOP
LED is turn-
ed on; then the channel
is opened,
and the Model
705 ad-
vances
to the next channel.
This sequence
is repeated
up to
the last channel
and then instead
of advancing
to the next
channel the Model 705 resets to the programmed
first channel
and the entire sequence
is repeated.
The scanning
sequence
can be stopped
at any programmed
channel
by simply pressing the START/STOP
button.
Press-
ing the START/STOP
button
during a continuous
scan stops
the scan at the presently
displayed
channel,
the channel
re-
mains closed, and turns off the START/STOP
LED. To start
the scanning
sequence
from the presently
displayed
channel,
press the
START/STOP
button.
The
START/STOP
LED
turns on when resuming
the scan,
To stop the scanning
sequence,
and reset the Model 705 to
the programmed
first channel,simply
press the RESET button.
Example
5 Temperature
Scan-With
the Model 7057 ther-
mocouple
plug in card, the Model 705 can scan separate temp-
erature points. The Model 7057 thermocouple
scanner card is
field installable in the Model 705 Scanner mainframe.
Since the
Model 7057 combines
the functions
of a thermocouple
scan-
ner and uniform
temperature
reference it is especially useful for
scanning
thermocouples.
The input
terminals
are #lO alloy-
copper set in an isothermal
block to minimize temperature
dif-
ferences.
A thermistor
sensor within
the isothermal
block
is
used with a bridge network
located on the Model 7057 to give
an indication
of the temperature
reference
or cold junction.
The temperature
of the heat sink is used to calculate
the cor-
rected thermocouple
output.
The output voltages
of each ther-
mocouple
must be converted
to temperature
(OC or OF) using
appropriate
thermocouple
tables or polynomial
equations.
In-
put and output connections
are made through
the rear panel of
the Model 705 Scanner mainframe
using #4 screw terminals
on
the Model 7057.
For complete
details concerning
the Model
7057 thermocouple
scanner card refer to the Model 7057 In-
struction
Manual.
In this example,
the Model 705 will be programmed
to scan
five separate temperatures.
Also, the following
parameters
will
be programmed
into the Model 705.
I. First channel
is 1.
2. Last channel
is 6.
3. Number
of poles is 2.
4. Interval time is three seconds,
5. Use the continuous
scan mode.
Use the following
procedure
to program the Model 705 for this
example:
1. If Example 4 is still set up, disconnect
it. Connect
the five
temperature
sources to the Model 7057 via the #4 screw
terminals
at the rear of the card.
2. Install the Model 7057 into the Model 705 Scanner
main-
frame.
Refer to paragraph
2.2.2 for scanner card installa-
tion instructions.
3. Select appropriate
line voltage
and turn on the Model 705.
Refer to paragraphs
2.2.1 and 2.3 respectively.
4. Press PRGM, 6, 2, ENTER. For complete
details concern-
ing front panel programs
refer to paragraph
2.6.
5. Press 0, 1, FIRST. (Programs
channel
1 as the first chan-
nel.) Channel
1 is used to monitor
the temperature
of the
input terminals
with the thermistor
bridge.
6. Press 0, 6,
LAST,
(Programs
channel
6 as the
last
channel.)
2-10

Advertisement

Table of Contents
loading

Table of Contents