NOTE
The alarm time is now set to 2:45 pm. The alarm
will go off everyday
at 2:45
pm outputting
the
negative
going TTL level pulse to the ALARM/
SERIAL
IN connector.
To verify that the alarm time was set to 2:45 pm by the pre-
ceding
example,
press PRGM, 7 and the display
shows that
the alarm time was set to 2:45 pm.
NOTE
Upon
power-up
or upon
activating
front
panel
Program
99, the Model 705 sets the alarm time
to 00.00.00.
When
the Alarm
time
is set for
00.00.00,
no pulse appears.
2.6.9 Program 8 Self Test
Program
8 is used to run a test on the ROM, RAM circuitry
and also to test the front panel
LED's
This test is also per-
formed
upon power-up.
If it is desired
to run the test without
powering
the unit down then up, Program
8 can be used. Use
the following
procedure
to activate
Program
8.
1. Press PRGM (selects program mode).
2. Press 8 (selects program 8).
Once step 2 is completed
the Model 705 performs
the follow-
ing sequence:
First the Model 705 turns on all the segments
in the display
as shown:
This is a display
test. The operator
can note inoperative
dis-
play segments
by comparing
the Model 705 display
with the
figure above.
In addition,
the push button and the TALK, LIS-
TEN, and REMOTE
indicators
turn on. All indicators
are on
simultaneously
if operating
correctly.
While all the LEDs are
on, the Model 705 is performing
a cyclic redundancy
check
(CRC) of the ROM circuitry. A digital self test of the RAM cir-
cuitry
is also performed
at this time. If the digital
self test
reveals
a problem
with the RAM circuitry
the Model 705 dis-
plays the following
for approximately
0.5 seconds.
I
1
I
After this display the Model 705 displays
the following:
,
I
FRIl
I
If the CRC
reveals
a problem
with the ROM
circuitry
the
Model
705
displays
the
following
for
approximately
0.5
seconds.
After this display the Model 705 displays
the following:
If the self test
and the
CRC
both
reveal
a problem
the
Model 705 displays
the RAM and ROM error messages
and
then displays
the fail message.
When either
test does
not reveal
a problem
the Model
705
displays
the following
message
which
indicates
both tests
have passed:
I
PRS5
I
I
2.6.10 Program 90 Stand Alone
I
Program
90 is used to program
a Model 705, out of the Mas-
ter and Slave programs
and into normal front panel operation.
1. Press PRGM. (Selects program mode.)
2.
Press 9, 0. (After pressing the 0 button the Model 705 dis-
plays the following two messages
and reverts to the channel
mode. The channel LED turns on and the channel previously
displayed before pressing PRGM is displayed.)
I
I
I
I
Program 90 takes the Model out of the Master condition
and
also out of the Slave condition.
Program 90 also set the pro-
grammed first channel to 1 and the programmed
last channel
to 1.
NOTE
After actuating
Program
90 cycle
power
to the
instrument
or activate
Program
99 in order to set
the instrument's
ALARM/SERIAL
IN connector
to alarm output.
2.6.11 Program 91 Master
Program
91 sets the Model
705 as the master
control
of a
daisy chain configuration.
The designated
master
Model
705
is the only unit in the configuration
that has front panel con-
trol. All the controls
on the front panel control
the master
unit
as well as the slave units of the configuration.
2-l 8