Keithley 705 Instruction Manual page 49

Table of Contents

Advertisement

9. Channel Save/Recall (I) and (Z}
The letter I is the command
to save the present relay set up
or recall the stored relay set up. Refer to paragraphs
26.5
and 2.6.6 (front panel Programs
4 and 5) for more informa-
tion concerning
saving and recalling
relay set ups. The for-
mat for the save/recall
command
is as follows:
II, 12, 13, I4 and 15 -
Save the present relay set up.
Zi , 22, 23, 24 and 25 -
Recall the stored relay set up.
Programming Example
-
Enter the following
statements
into
the HP-85 to program the Model 705 to save the present relay
set up.
REMOTE 717 (END LINE)
OUTPUT 717; "11X" (END LINE)
When END LINE is pressed
the second time, the Model 705
saves the present relay set up. To recall the stored relay set up
enter the following statements
into the HP-85
REMOTE 717 (END LINE)
OUTPUT 717; "Zl X" (END LINE)
When END LINE is pressed
the second time, the Model 705
recalls the stored relay set up.
Programming Example
-
Use the following program to store a
relay set up of channels
1 through 5 with the programmed
first
channel
being
channel
1, and the programmed
last channel
being channel 5. The odd channels
are programmed
closed and
the even channels are programmed
open. Type the following pro-
gram into the HP-85:
PROGRAM
COMMENTS
10
20
30
40
50
60
70
80
REMOTE 717
(END LINE)
CLEAR 7 (END LINE)
OUTPUT 717; "Bl Fl
ClX"
(END LINE)
OUTPUT 717; "B2N2X"
(END LINE)
OUTPUT 717, "B3C3X"
(END LINE)
OUTPUT 717; "B4N4X"
(END LINE)
OUTPUT 717;
"B5L5C5llX"
(END
LINE)
OUTPUT 717; "RX"
90
END
-
Sets the Model 705 for the
remote mode.
Clears the Model 705.
Displays, closes and sets
channel 1 as the first channel.
Displays and closes channel 2.
Displays and closes channel 3.
Displays and opens channel 4.
Displays, closes, sets channel 5
as last channel and saves relay
set up.
Opens all channels and displays
the first channel.
Press the Run key on the HP-85 to run the program. After the
run key is activated the Model 705 displays the reset condition
Use the following programming
example to recall and display the
saved relay set up.
PROGRAM
COMMENTS
Output 717; "Zlx"
(END LINE)
Recalls saved relay set up,
This program
and the previous
nine step program
saved and
recalled one relay set up. The Model 705 is capable
of saving
and recalling five set ups. The commands
to save and recall the
relay set ups are as follows:
I1 saves a set up and Zl recalls the II set up.
12 saves a set up and 22 recalls the I2 set up.
13 saves a set up and 23 recalls the I3 set up.
I4 saves a set up and 24 recalls the I4 set up.
I5 saves a set up and
25
recalls the I5 set up.
It's possible to save a relay set up in any of the I commands with-
out having any of the other save commands
programmed
to save
a set up. For example,
12 can save a set up without having a set
up saved by II, 13, I4 or 15.
10. ROM and LEDTest (J)
The letter J is the command that programs the Model 705 to
go through many of the testing routines that are automati-
cally performed
upon power-up.
When the self test com-
mand is sent, the following actions are performed:
1. ROMTest
2.
LED Test (All front panel LEDs turn on for a short period
of time.)
3.
If the tests did not find anything wrong, then the self test
(J) byte in the status word is set to one. If the test failed
then the self test byte (J) is set to two.
The self test command
is implemented
with the following
command sequence:
OUTPUT 717; "JOX" (END LINE)
NOTES
1.
Other
operating
parameters
are not affected
by the JO
command.
2.
RAM is not tested by the JO command
but is tested upon
'power-up.
11, EOI (End or Identify) (K)
The letter K is the command that programs the Model 705 to
either send or not send EOI on the last byte out on a multi-
ple byte sequence.
EOI response of the instrument
may be
programmed
as follows:
KO -
Send EOI during last byte.
Kl -
Send no EOI.
Upon
power-up,
or after
receiving
an
SDC
or
DCL,
the
Model 705 will be in the KO mode.
Programming Example
-The
EOI will be suppressed
with the
following statement sequence:
REMOTE 717 (END LINE)
OUTPUT 717; "Ki X" (END LINE)
Note that the HP-85 does not rely on EOI to mark the end of the
data transmission.
Some controllers,
however, may require that
EOI be present at the end of transmission.
3-18

Advertisement

Table of Contents
loading

Table of Contents